Valhalla Scientific 2110A Overview
Models 2110A and 2111A provide a fast and convenient method of determining product efficiency, power factor, and true RMS current draw. Phase angle relationships may be calculated through manipulation of the displayed quantities.
Specified accuracies are valid for a period of 1 year from the date of calibration at 25°C ±5°C, following a 30 minute warm-up.
Voltage - AC+DC, DC Coupled DC and 40Hz - 5kHz: ±0.1% of reading ±6 digits
5kHz - 15kHz: ±0.5% of reading ±6 digits
15kHz - 20kHz: ±0.75% of reading ±6 digits
(Useable above 20kHz to 50kHz with typically an additional 1% error per 10kHz)
Current - AC+DC, DC Coupled
DC and 40Hz - 5kHz: ±0.1% of reading ±6 digits
5kHz - 15kHz: ±0.5% of reading ±6 digits
15kHz - 20kHz: ±0.75% of reading ±6 digits
(Useable above 20kHz to 50kHz with typically an additional 1% error per 10kHz)